Gradient high-performance thin-layer chromatography for characterizing complex hydrocarbon-containing products

3.1 Classic SARA and SARA expanded by gradient

Figure 1A and B show a classic SARA chromatogram of a Canadian bitumen, using a sequence of several elution solvents in order of increasing polarity. In this case, the AMD instrument was used as a conventional developing chamber. As detailed in Table 1, saturates (Sat) were first separated (with C7 at 60 mm m.d.), then aromatics (Ar, with DCM at 40 mm m.d.), resins (Res, with THF at 20 mm), and asphaltenes (Asph, non-developed peak at the application point). Sat were detected after plate impregnation by berberine-induced fluorescence (λexc = 365 nm; λem > 400 nm) (Fig. 1B). Ar, Res, and Asph were detected by UV at 273 nm, the wavelength that provided maximum absorption for each of these groups (Fig. 1A).

Fig. 1figure 1

Classic SARA chromatogram of a Canadian bitumen: A UV 273 nm and B berberine-induced fluorescence 365 nm. A 23-step AMD gradient of a Canadian bitumen: C UV 273 nm and D FDIC 365 nm

When a 23-step AMD gradient based on the same eluents (THF‒DCM‒C7, see conditions in Table 2) was applied on the same sample, the resulting chromatogram is shown in Fig. 1C and D. As we can observe, the simplicity of the classical SARA chromatogram (Fig. 1A) is deceptive. The 23-stage SARA (Fig. 1C) gives us a clearer idea of the complexity of the sample.

AMD operates in successive elemental elution steps (of several millimeters each), which are performed with decreasing solvent strength, in our case THF > DCM > C7, in combination with increasing developing distance. The composition of the mobile phase in each step does not have to be the same and usually is of lower polarity that than of the previous one, adjusted by mixing of different proportions of the three selected solvents. The stepwise gradient formed allows narrow peaks and more efficient separations to be obtained in a highly reproducible way (± 0.45 mm). Repetitions of elemental steps (twice) were performed to provide band refocusing.

3.2 Gradient optimization: from 23-step SARA to 20-step SARA

Figure 2 shows the 23-step chromatogram (Table 2) of a hydrorefined vacuum gas oil at different UV wavelengths and FDIC. AMD chromatographic zones can be related to peaks from hydrocarbon groups in classical SARA. The peak in the gradient chromatogram appearing in the 100 vol% THF development zone (up to 14 mm m.d.; one step) can be considered as asphaltenes (non-developed by THF). The peaks eluted in the 18‒58 mm m.d. zone (10 steps, from 100 vol% DCM to 20:80, V/V, DCM‒heptane), correspond to resins. This zone is particularly well resolved using the described gradient conditions and may provide more useful information than that obtained using classical methods. The peaks eluted in the 58‒83 mm zone, from 15:85, V/V, DCM‒heptane to 100 vol% n-heptane, are usually considered aromatics and saturates.

Fig. 2figure 2

A Chromatogram of a vacuum gas oil at different wavelengths: UV 273 nm (–), UV 228 nm (

figure a), UV 201 nm (figure b), and FDIC 365 nm (figure c) under the AMD conditions (Table 2, 23 steps), and B under the AMD conditions (Table 3, 20 steps); C zoom of apolar zone (55‒85 mm)

An inefficient zone with regard to separation (57–67 mm, m.d.) was identified in this chromatogram. As in general AMD practice, an unresolved chromatogram can be superimposed with the corresponding diagram of the gradient employed, changes in solvent composition can be made, and a number of steps can be done to improve gradient separation. In this way, the gradient conditions were modified to 20 steps (see Fig. 2B and conditions in Table 3) to optimize separation zones on the plate. Thus, separation can be fine-tuned and amplified for each segment of the chromatogram. In our case, the separation in the apolar zone was improved without losing resolution in the asphaltenes‒resins zone and by removing the inefficient plate zone. To do this, steps 12 to 17, which included DCM‒C7 gradient, were removed, and seven final steps of C7 (100%) were added. In total, the sequence went from having 23 to 20 steps. Details of both gradients are described in Tables 2 and 3.

The saving of three steps in the chromatographic development involved going from 5 h and 20 min to 5 h of analysis time. Regardless of the final number of steps and despite the duration of these analyses, the number of substances processed per plate compensates the time invested with regard to other chromatographic techniques. The AMD instrument can be left programmed to perform the analysis overnight in automatic mode.

3.3 Heavy product characterization using optimized 20-step THF‒DCM‒C7 SARA gradient

Figure 2B shows the chromatogram of the mentioned hydrorefined vacuum gas oil under these AMD conditions at different UV wavelengths and FDIC. On-silica UV spectra of standards are similar to those obtained in dissolution and provide useful information for selecting densitometric conditions. UV spectra of several standards are shown in the Supplementary Information.

The most polar and heaviest compounds (Asph) remain at the application point (m.d. = 10 mm) and are directly detected on the plate by UV‒densitometry. Resins zone (18‒58 mm m.d. zone, 10 steps, from 100 vol% DCM to 20:80, V/V, DCM‒C7) is particularly well resolved using the described gradient conditions. As we will show below, the role of THF is crucial to achieve an adequate resolution of this zone.

Peaks eluted in the 58‒83 mm zone, seven steps at 100 vol% C7, are Aromatics and Saturates. This zone is especially interesting and rich in aromatic and saturated compounds in hydro-refining products such as this one, as can be seen by comparing the profile of this product with that of a bitumen in Fig. 1B.

Figure 2C shows a detailed view of different chemical families in this zone of the chromatogram where several peaks are coeluted. Identification of component groups has been possible by using selective detection, direct recording of UV spectrum of each separated peak, and comparing UV spectra of peaks with those of standards that were eluted on the plate under the same gradient conditions.

The peak eluted at 83.3 mm with the front of solvent is due to saturated hydrocarbons. This peak has no UV response and is detected by FDIC‒berberine [5]. An ion-induced dipole interaction between n-alkanes and berberine cation is responsible for an increase in berberine emission in the presence of alkanes [5]. The shoulder at 81 mm in the FDIC‒berberine chromatogram, which coelutes with saturated hydrocarbons and has a response in the presence of berberine, may be attributed to naphthenes. This shoulder is encompassed in the peak eluted at 80.3 mm, which has been selectively detected by UV at 201 nm. UV spectrum suggests that this peak consists of naphthenic and naphthene‒aromatic structures.

The inspection of the chromatogram obtained by UV detection at 228 nm, and the UV spectrum of the peak eluted at 78.2 mm, are coherent with the presence of alkyl-aromatic structures in this peak. Likewise, the chromatogram obtained by UV at 273 nm together with UV spectra of peaks eluted at 73.2, 68.5, and 64.3 mm suggest the presence of aromatic structures, probably with different degrees of condensation. Migration distances of standards under these conditions and the corresponding UV spectra are given in Supplementary Information.

Therefore, this optimized 20-step THF‒DCM‒C7 gradient can be used for profiling a variety of samples that covers the whole range of petroleum heavy products.

3.4 Applications of expanded SARA gradient

The application of multi-step gradient sequence to different heavy products provides qualitatively different chromatograms among products. Although this separation does not provide a complete molecular resolution of heavy petroleum products, the resulting profiles adequately represent the complexity of the analyte, and can provide a useful information for evaluating the effect of process variables on products. We show here some specific applications to different heavy products.

3.4.1 Bitumen

Bitumen is a black, viscous mixture of hydrocarbons obtained naturally or as a residue from petroleum distillation [18]. It is is too thick to be pumped from the ground or through pipelines.

Different qualitative profiles were obtained by application of 20-step expanded SARA gradient (THF‒DCM‒C7) in the function of their origin, regardless of their distillation cut temperature (DCT; in degrees celsius, °C). Figure 3 shows different profiles for bitumen from Iran (Norooz, DCT at +420, +450, and +490 °C), Gabon (Oguendjo/Olende, DCT at +495, +520, and +560 °C), and Cameroon (Lokele, DCT at +440, +470, and +490 °C).

Fig. 3figure 3

A Bitumen Norooz +420 °C, B Bitumen Norooz +450 °C, C Bitumen Norooz +490 °C, D Oguendjo/Olende +495 °C, E Oguendjo/Olende +520 °C, F Oguendjo/Olende +560 °C, G Lokele +440 °C, H Lokele +470 °C, and I Lokele +490 °C. 20-Step AMD gradient, UV 273 nm

It seems possible to obtain fingerprints for bitumen characterization by HPTLC in a similar way to those obtained in the case of characterization of plants, therefore relating bitumen composition and properties.

3.4.2 Asphaltenes

The structure of asphaltenes is still a challenge for analytical chemistry. From the point of view of solubility, asphaltenes are insoluble in C7 and soluble in toluene. This is the way to obtain them gravimetrically. Recent research suggests that they are “hand” shaped entities, with a central core consisting of a hydro/polyaromatic system (for example, around seven cores with a few N and S heteroatoms) with peripheral alkanes [19]. The molecular masses of the asphaltenes are between 300 and 1400 Da, with the majority of species between 400 and 800 Da.

Chromatographically, asphaltenes are associated with the undeveloped peak that remains at the point of application (in Iatroscan and in HPTLC) [13], or with the part that remains irreversibly retained in the stationary phase of chromatographic columns (in the case of HPLC and open column chromatography, as in the D2007 standard [6]).

When Asph are gravimetrically isolated they tend to “catch” some Res, structures that resemble those of Asph with a lower molecular weight and that are strongly attracted by them. Figure 4 shows that the 23-step gradient reveals different “resinic part” in two different types of Asph. This gradient is also useful to detect trapped Sat in Asph structures (not shown).

Fig. 4figure 4

A, B Asphaltenes from different origins. 23-Step AMD gradient, UV 273 nm

3.4.3 Hydrorefining products

Hydrorefining is a process for improving the quality of a petroleum product by hydrogenating in the presence of a catalyst at a temperature below that at which decomposition takes place [20]. This improves the H/C ratio of the product. Thus, a consequence of hydrorefining, deasphalting is observed (a decrease in the Asph peak), and also an increase in the Saturates fraction. This can be observed in Fig. 5 where the 23-step gradient chromatograms of a heavy Canadian oil and its derived hydrorefined product are shown.

Fig. 5figure 5

The 23-step AMD gradient of heavy Canadian oil: A UV 273 nm, B FDIC 365 nm. Derived hydrorefined of heavy Canadian product: C UV 273 nm, D FDIC 365 nm

3.4.4 Base oils

Base oils, produced from refining crude oil, are used in the formulation of lubricant products. They are complex mixtures containing hydrocarbon families of very different polarity, such as paraffins, isoparaffins, naphthenes (cycloparaffins), hydroaromatics, aromatics with different condensation degrees, and a variety of S-, O-, and N-heterocyclic structures, with molecular weight distributions ranging from 300 to 600 Da, although in general, they are asphaltene-free and have a lower polarity than other heavy products [21]. Base oils only have compounds in the apolar zone of the 20-step chromatogram (Fig. 6).

Fig. 6figure 6

The 20-Step AMD gradient of base oil: (–): UV 273 nm, (

figure d): UV 228 nm, (figure e): UV 201 nm (figure f): FDIC 365 nm

Under the 20-step AMD conditions, it is possible to distinguish the families of naphthenes, naphthenoaromatics, and alkylaromatics between 75 and 83 mm, with a degree of detail. This zone corresponds to the last seven stages of the gradient, with a composition 100% C7, according to Table 3. The families in base oil D were detected by UV detection at 201, 228, and 273 nm and FDIC, and UV spectra of the separate peaks were consistent with those of the model compounds studied (Supplementary Information).

Although aromatic compounds present some coelution in this zone of chromatogram, a selective UV detection at different wavelengths and the UV spectra of obtained peaks allowed us to identify several types of compounds in base oil D: saturates (m.d. = 84.3 mm), naphthenes (m.d. = 82.5 mm, a shoulder in saturates peak), heavy naphthene/naphthene-aromatics (m.d. = 80.4 mm), and alkyl-aromatics (m.d. = 78.3 mm). In the case of the peak at 80.4 mm, there would be a greater contribution from heavy naphthenes than from aromatics, since little absorbance is observed at 273 nm.

3.5 The role of THF in the separation

We have found that THF plays a crucial role in the degree of resolution of extended separations of heavy petroleum products, and particularly in the good separation obtained in the resins zone.

When THF is used either as a plate precleaning solvent in preconditioning step, or as a mobile phase in the gradient, good resolutions are obtained in the resulting chromatograms. On the contrary, when THF is neither used as a washing solvent nor as a mobile phase, profiles completely lose their resolution.

For evaluating the effect of THF as a mobile phase in the gradient, a 20-step DCM‒C7 gradient was used (Table 4) to compare with the gradient containing THF (Table 3). Therefore, the first two steps eluted with THF were changed to two steps eluted with DCM. Moreover, the effect of DCM as plate preconditioning solvent was compared with that of THF. The sample studied was a vacuum gas oil.

Figure 7 summarizes the results obtained for DCM‒C7 gradient. When THF was not present in the chromatographic system, i.e., using DCM‒C7 gradient, and DCM was the washing solvent, resolved peaks disappeared, and resins compounds were eluted to the apolar part of the chromatogram (Fig. 7A). Using this same gradient, resolved peaks in the resins zone are present when THF has been used in the preconditioning, washing step (Fig. 7B).

Fig. 7figure 7

Vacuum gas oil, 20-step DCM‒C7 gradient: A DCM as washing solvent, B THF as washing solvent, C THF as washing solvent removed by plate heating for 30 min at 105 °C, D without washing step

Figure 7C shows that a low-resolution chromatogram, similar to that of Fig. 7A, is obtained when THF is used in the washing step but is removed from the plate heating (heating for 30 min at 105 °C) before sample application. Likewise, Fig. 7D gives an idea of the importance of the precleaning operation. When no washing step is included, an intense peak of impurities is eluted by the gradient at a migration distance of 72 mm.

Results obtained from the 20-step THF‒DCM‒C7 gradient from the same vacuum oil sample are given in Fig. 8A–D. In all cases, resolved chromatographic profiles were obtained, with an adequate degree of resolution for the compounds eluted in the resins zone when THF was present either in the gradient mobile phase or as a washing solvent.

Fig. 8figure 8

Vacuum gas oil/hydrorefined product, 20-step THF‒DCM‒n-heptane gradient: A DCM as washing solvent, B THF as washing solvent, C THF as washing solvent (plate heating for 30 min at 105 °C), D without washing step

The effect of washing with DCM (Fig. 8A) or THF (Fig. 8B) gives a different distribution proportion of developed compounds. Likewise, the removal of THF by heating (Fig. 8C) shows a chromatogram similar to that of Fig. 8A (with DCM as the precleaning solvent). Moreover, Fig. 8D shows that using the plate with previous cleaning provided different repartition and proportion of resins onto the silica gel plate.

It is well known that THF adsorbs on silica gel surfaces, and that distinct mechanisms of adsorption (nonspecific, specific, molecular, and chemical adsorption) were observed [22]. Likewise, it has been found that adsorption isotherms of polar solvents on silica gel does fit abi-layer adsorption equation rather than the mono-layer one [23]. According to Scott [23], such layers can impede the interaction of the solute directly with the stationary phase as in sorption.

A combined and interactive mechanism of sorption and displacement was proposed by Scott when the silica surface was covered with a layer of molecules from one or more different solvents (Fig. 9).

Fig. 9figure 9

Mechanism of sorption and displacement of THF on silica gel surface (

figure g: Solute and figure h: THF): a solute interacting with first layer of THF by sorption, b solute interacting with second layer of THF by sorption, c solute interacting with first THF layer by displacing THF from second layer, d solute interacting with silica surface by displacing THF from second layer

In our case, THF is the solvent that interacts directly with the silica surface in the precleaning step, and consequently constitutes the first layer. Thus, elution began using pure THF, and the next layer was a second layer of THF. Thus, a monolayer or bilayer of THF may be formed on the silica surface with which the solute interacts by sorption (mechanisms a and b). In general, the stronger the polarity of the solute, the more likely it is to interact with the surface by displacement (mechanism c), even to the extent of displacing both layers of THF (mechanism d). This influences the separation of the resins in the petroleum products studied.