Mechanical QT and JT intervals by M-mode echocardiography: An extrapolation from the concurrent electrocardiographic tracings

   Abstract 


Background : Congenital long QT syndrome (CLQTS) is a life-threatening ion channelopathy leading to syncope and sudden death. Early diagnosis during the prenatal period and timely intervention can prevent sudden cardiac death and catastrophic consequences of this genetic ion channelopathy. Fetal magnetocardiography and fetal electrocardiography (ECG) enable the measurement of fetal QT and JT intervals, but their inherently technically challenging and/or resource-intensiveness nature preclude their routine clinical application. On the other hand, the high-temporal resolution of M-mode echocardiography makes it a well-suited and widely available modality for the measurement of cardiac events.
Aims and Objectives : We aimed to investigate the mechanical counterparts of the electrical QT and JT intervals on M-mode echocardiographic images of the tricuspid, mitral and aortic valves, and aortic wall.
Methods : We performed a prospective study on consecutive children referred to the outpatient pediatric cardiology clinic at a tertiary children's hospital. We defined M-mode echocardiographic landmark points on tracings of tricuspid annular planar systolic excursion, mitral and aortic valves, and aortic wall with simultaneous electrocardiographic recording. We measured the mean±SD of the absolute time difference and RR-adjusted time difference in cases with non-coincident ECG events and echocardiographic landmarks.
Results : Fifty healthy children were enrolled in the study. In 47 (94%) out of the 50 children, Q was coincident with the starting point of the tricuspid annular plane systolic excursion. In all children, the Q was coincident with the mid-point of the A-C line of the mitral valve. In 38 (76%) cases, there was a bump on the anterior wall of the aortic root immediately before the change in the slope of the aortic wall. This was coincident with the Q wave in 100% of cases. In all cases, the J point coincided with the point of acceleration of velocity on TAPSE. In all children, the J point coincided with the initial maximal opening of the aortic cusps. The end of the T wave occurred coincident with the peak of the tricuspid annular planar systolic excursion in 47 children (94%). In 48 children (96%), the end of the T wave coincided with the aortic cusps' closure point.
Conclusions : Based on our findings, we propose to measure the averaged mechanical QT and JT intervals by using an angled M-mode tracing of the aortic and mitral valve in five consecutive beats in the parasternal long-axis view. This is the first study on mechanical QT and JT intervals in healthy children. The study opens the horizons into the in-utero diagnosis of congenital long QT syndrome by measuring fetal QT and JT intervals using the widely available M-mode echocardiography

Keywords: Congenital long QT syndrome, M-mode echocardiography, mechanical JT interval, mechanical QT interval

How to cite this article:
Rad EM, Ilali HM, Majnoon MT, Zeinaloo A. Mechanical QT and JT intervals by M-mode echocardiography: An extrapolation from the concurrent electrocardiographic tracings. Ann Pediatr Card 2022;15:364-73
How to cite this URL:
Rad EM, Ilali HM, Majnoon MT, Zeinaloo A. Mechanical QT and JT intervals by M-mode echocardiography: An extrapolation from the concurrent electrocardiographic tracings. Ann Pediatr Card [serial online] 2022 [cited 2023 Jan 7];15:364-73. Available from: 
https://www.annalspc.com/text.asp?2022/15/4/364/367286    Introduction Top

Congenital long QT syndrome (CLQTS) is a life-threatening ion channelopathy leading to syncope and sudden death.[1],[2],[3],[4],[5] This channelopathy is frequently undiagnosed with an underestimated prevalence of 1/3000–10,000. Given the presence of cardiocerebral channelopathy, the timely diagnosis of this inherited abnormality becomes paramount.[6] Early diagnosis during the prenatal period and timely intervention can prevent sudden cardiac death and catastrophic consequences of this genetic ion channelopathy.[5] Although specific features such as bradycardia, complete heart block, tachydysrhythmia, including ventricular tachycardia, and evidence of heart failure have been reported with higher frequency in fetuses with long QT syndrome, robust tools are essential for reliable prenatal diagnosis of this potentially life-threatening inherited arrhythmia.[7],[8],[9],[10],[11],[12]

Mechanical PR interval, measured from the onset of the A-wave of the mitral valve to the onset of the flow of the aortic valve, has been introduced and validated as a surrogate for electrical PR interval in the fetal period.[13],[14],[15] Moreover, despite the emergence of advanced echocardiographic modalities, M-mode echocardiography is still a time-honored modality because of its superior temporal resolution, which enables recording of the cardiac events at a frame rate of 1000 per second.[16]

To the best of the authors' knowledge, there is no reported study on echocardiographic estimation of mechanical QT and JT to date. Therefore, inspired by the concept of mechanical PR and considering the advantage of the high temporal resolution of widely available M-mode echocardiography, we aimed to delineate the corresponding mechanical counterparts of the electrical QT and JT intervals on M-mode echocardiographic images of the tricuspid, mitral and aortic valves, and aortic wall by using the concurrent electrocardiographic (ECG) tracing on the monitor of the echocardiographic machine. Similarly, we defined the echocardiographic landmarks of the JT interval in our study because this interval is a more accurate measure of ventricular repolarization, particularly in patients with ventricular conduction abnormalities.[17],[18]

   Methods Top

Study design and study population

We performed a prospective study on consecutive children referred to the outpatient pediatric cardiology clinic of the Children's Medical Center for evaluation of cardiac murmur. After a complete cardiac examination and comprehensive color Doppler two-dimensional echocardiographic examination, children with normal hearts and informed consent of their parents or guardians were enrolled in the study. Patients with poor acoustic windows were excluded.

Estimation of minimum sample size

The sample size was estimated using G*Power software (version 3.1.9.7, Germany).[19] At a power level of 95% and a significance level of 0.05%, a minimum sample size of 45 was estimated.

Echocardiographic examination

Echocardiography was performed using Philips Affiniti 50 C and probes S8-3 and S4-2 (Philips Healthcare, 3000 Minuteman Road, Andover, MA 01810, USA). Color tissue Doppler imaging of tricuspid annular planar systolic excursion (TAPSE), M-mode of the mitral and aortic valve, and aortic wall were obtained according to recommendations of the American Society of Echocardiography.[20] The simultaneous ECG recording was performed in all cases. All echocardiographic examinations were saved, and echocardiographic counterparts of Q, J, and end of T-wave were marked on recorded images of TAPSE, mitral and aortic valves, and the aortic wall on parasternal long-axis view. The sweep speed of the electrocardiography was adjusted according to the heart rate of the patient. We chose a channel that contained a recognizable Q- or q-wave or an initial r.

We chose four M-mode parameters: longitudinal M-mode parameter of TAPSE recorded by color tissue Doppler imaging, M-mode tracings of mitral and aortic valves, and the aortic wall on parasternal long-axis view.

Initially, a pilot study was performed to determine the most common landmarks on the M-mode tracings of the TAPSE, mitral and aortic valves, and the aortic wall. Consequently, echocardiographic landmark points were defined as described below on TAPSE and M-mode images of mitral and aortic valves and aortic walls [Figure 1].

Figure 1: Echocardiographic landmarks. Red, orange, and blue bullets represent Q, J, and end of the T-wave, respectively. (a) ECG landmarks, (b) Echocardiographic landmarks on M-mode of mitral valve obtained in parasternal long-axis view, (c) Echocardiographic landmarks on M-mode of the aortic valve and anterior aortic wall obtained in parasternal long-axis view, (d) Echocardiographic landmarks on the TAPSE obtained in apical four-chamber view, (e) Extrapolation of Q, J, and end of T-wave on the echocardiographic landmarks on the tracing of TAPSE: Q coincides with the start of excursion, J occurs at the time of velocity acceleration, and the end of T-wave is coincident with the peak of systolic excursion, TAPSE: Tricuspid annular planar systolic excursion, ECG: Electrocardiographic

Click here to view

Tricuspid annular plane systolic excursion

The beginning point of the systolic excursionPoint of change in velocity (acceleration of velocity) in systolic excursion marked by a change in color of the recoding from red to bright red or orangePoint of peak systolic excursion.

M-mode of the mitral valve (on parasternal long-axis view)

Mid-point of A-C linePoint of closure of the leaflets (C)Point of the beginning of the opening of the leaflets (D)

M-mode of the aortic valve (on parasternal long-axis view)

Point of the beginning of the opening of the leafletsInitial complete separation of the leafletPoint of closing of the leaflets.

M-mode of the anterior aortic wall (on parasternal long-axis view)

Early systolic bump on the anterior aortic wallThe nadir on the anterior aortic wallAortic wall peak anterior motion.

Measurement of the time difference between the echocardiographic landmark and the electrocardiographic event

In patients in whom some beats or all were noncoincident with the predefined echocardiographic landmarks, the time differences between the echocardiographic landmark and the ECG event were measured and averaged [Figure 2].

Figure 2: The extrapolation of the J point on the anterior aortic wall. In the first, third, and fourth beats, J occurred 0.09, 0.02, and 0.05 s later than the nadir on the anterior aortic wall, respectively. The average of these time differences is 0.05 s

Click here to view

All measurements were carried out by a senior consultant (EMR). The echocardiographic studies of 17 children were repeated at an interval to assess the intra-observer variability of the measurements. Cohen's kappa and intra-class correlation coefficient (ICC) were calculated for categorical and continuous variables, respectively.

Statistical analysis

Data were stated as number and frequency. We used IBM SPSS Statistics for Windows, version 25 (IBM Corp., Armonk, NY, USA) for statistical analysis. Shapiro–Wilk test was used to test the normality of the distribution of the variables. In cases where the ECG events were not coincident with the defined landmarks, we calculated the Pearson's correlation coefficient between the QT and JT intervals on ECG tracings on the echocardiographic screen and the corresponding intervals based on the echocardiographic landmarks. Furthermore, we presented the mean ± standard deviation (SD) of the absolute time difference and the RR-adjusted time difference (i.e., time difference divided by the RR interval) between ECG events and echocardiographic landmarks in these cases.

Ethical considerations

We obtained informed consent from the children's parents or guardians and the participants, if applicable. We followed the principles stated in the Helsinki Declaration on ethical principles for medical research involving human subjects.[21] The Institutional Committee approved this study for Ethics in Biomedical Research of the Medical School of Tehran University of Medical Sciences (ID: IR.TUMS.CHMC.REC.1399.190).

   Results Top

The basic characteristics of the study population are shown in [Table 1]. Fifty healthy children were enrolled in the study.

Extrapolation of Q of electrocardiographic on M-mode echocardiography of tricuspid annular plane systolic excursion, mitral and aortic valves, and anterior aortic wall

Tricuspid annular plane systolic excursion

In 47 (94%) out of the 50 children, Q was coincident with the starting point of the TAPSE. Only in three (6%) children, Q occurred slightly sooner than this point. The mean ± SD time difference between the Q on ECG and the starting point of TAPSE in noncoincident cases was 0.036 ± 0.011 s [Table 2].

Table 2: Mechanical counterparts of three points of Q, J, and end of T-wave on four M-mode tracings: Tricuspid annular planar systolic excursion, mitral valve, aortic valve, and aortic wall in 50 healthy children

Click here to view

M-mode of mitral valve

In all children, the Q was coincident with the mid-point of the A-C line of the mitral valve [Figure 3], [Figure 4], [Figure 5].

Figure 3: (a and b) Schematic illustration showing extrapolation of Q, J, and end of T-wave on the echocardiographic landmarks on the M-mode tracing of the mitral valve obtained in the parasternal long-axis view. The red, orange, and blue bullets indicate Q, J, and the end of the T-wave, respectively. Q was coincident with the mid-point of the AC line in all cases. J and end of T-wave occurred on C and D points in 86% and 92% of the study population, respectively. The percentages in the box in the right lower corner show the coincidence of Q, J, and end of T-wave on ECG with these echocardiographic landmarks in this study. (c) M-mode tracing of the mitral valve obtained from the parasternal long-axis view shows the Q, J, and end of the T-wave. In this case, the J point occurs after the echocardiographic landmark (i.e., the C point), ECG: Electrocardiographic

Click here to view

Figure 4: (a) Extrapolation of Q, J, and the end of T-wave on the echocardiographic landmarks of the M-mode tracing of the anterior aortic wall recorded from the parasternal long-axis view, (b) Extrapolation of the J point on the M-mode of the anterior aortic wall, (c) Extrapolation of Q, J, and end of T-wave on the echocardiographic landmarks of the color tissue Doppler M-mode imaging of the anterior aortic wall recorded from the parasternal long-axis view, (d) A magnified view of the QJ bump is shown

Click here to view

Figure 5: The proposed echocardiographic landmarks for measurement of mechanical QT and JT intervals on M-mode echocardiography

Click here to view

M-mode of aortic valve

In 29 (58%) cases, Q occurred precisely coincident with the opening point of the aortic valve. However, in 31 children, it occurred slightly before the opening of the aortic valve, with a mean ± SD of time difference of 0.047 ± 0.011 [Figure 1].

M-mode of anterior aortic wall

In 38 (76%) cases, there was a bump on the anterior wall of the aortic root immediately before the change in the slope of the aortic wall. This bump was coincident with the Q-wave in 100% of cases [Figure 4]. However, this bump was absent in 12 (24%) children.

Extrapolation of J of electrocardiographic on M-mode echocardiography of tricuspid annular plane and mitral and aortic valves

Tricuspid annular plane systolic excursion

In all cases, the J point coincided with the point of acceleration of velocity [Figure 1]d and [Figure 1]e.

M-mode of mitral valve

J point was coincident with the C point in 43 (86%) of children. Notwithstanding, in seven cases (14%), the J point happened slightly after the C point at a mean ± SD time delay of 0.037 ± 0.004 s.

M-mode of aortic valve

In all children, the J point coincided with the initial maximal opening of the aortic cusps.

M-mode of the anterior aortic wall

In 40 (80%) children, J point occurred at the nadir on the anterior aortic wall on M-mode echocardiography in the parasternal long-axis view. In the remainder, it happened after the nadir on the aortic wall. In the presence of the bump, this nadir point was also coincident with the end of the bump on the anterior aortic wall. The beginning of this bump reflected Q and the end of the J point [Figure 4].

Extrapolation of end of T (T end) of electrocardiographic on M-mode echocardiography of tricuspid annular plane and mitral and aortic valves

Tricuspid annular plane systolic excursion

The end of the T-wave occurred coincident with the peak of the TAPSE in 47 children (94%). In three cases, the end of the T-wave happened earlier at a mean ± SD of time difference of 0.050 ± 0.010 [Figure 1]e.

M-mode of mitral valve

In 46 (92%) children, the end of the T-wave was coincident with the D point, and in only four cases (8%), it occurred 0.047 ± 0.018 s earlier [Figure 1]b and [Figure 1]c.

M-mode of aortic valve

In 48 (96%) children, the end of the T-wave coincided with the aortic cusps' closure point. In only two cases (4%), the end of the T-wave occurred later at a delay of 0.03 and 0.04 s, respectively.

M-mode of the anterior aortic wall

In 30 (60%) children, the end of the T-wave was simultaneous with the peak anterior motion of the anterior wall of the aortic wall on M-mode imaging in parasternal long-axis view [Figure 4]d.

Time difference with echocardiographic landmark points in noncoincident cases

The comparison of the values of electrocardiographic-based and predefined echocardiographic landmark-based QT and JT intervals on tricuspid annular plane systolic excursion, M-mode of mitral and aortic valves, and anterior aortic wall is depicted in [Table 3].

Table 3. Comparison of the values of ECG-based and pre-defined Echocardiographic Landmark-based QT and JT intervals on TAPSE, M-mode of mitral and aortic valves and anterior aortic wall

Click here to view

There was excellent intra-observer reproducibility between the pairs of the repeated measures and observations on categorical and continuous variables (ICC and Cohen's kappa coefficient ≥0.99, respectively).

   Discussion Top

This study is the first investigation to define mechanical QT and JT intervals using the M-mode echocardiography [Table 4].

Table 4: Summary of M-mode echocardiographic corresponding points of Q, J, and T with their respective coincidence percentages (M-mode and electrocardiographic landmarks) and permutation combinations for calculation of mechanical QT and JT intervals

Click here to view

Corresponding points of Q, J, and end of T-wave on M-mode echocardiography

Q-wave

In all cases, the Q-wave coincided with the onset of the bump on the anterior aortic wall. Since the beginning of this bump was coincident with Q and the end of that with J, we called this QJ bump. Similarly, the Q-wave and mid-point of the A-C line of the mitral valve co-occurred in all cases. Of note, given the relationship between the A-C duration and PR interval, this finding applies to children with normal PR intervals.

The origin of the QJ bump is not elucidated. Tan and Tsang noted the exquisite motion and function of the aortic wall.[26] The anterior wall of the aorta is in continuity with the interventricular septum. We speculate this right-ventricular inclined bump may correspond with the initial vector of depolarization that depolarizes the ventricular septum from left to right and from apex to base and produces an initial q in lead V6 and an r in lead V1.[27] This speculation needs to be confirmed.

J point

The J point coincided with the point of change in velocity or velocity acceleration point on TAPSE and the initial separation of the aortic valve leaflets in all cases.

End of the T-wave

Closure point of the aortic valve corresponded to the end of the T-wave in 96% of cases.

Q, J, and end of T-wave noncoincident with the echocardiographic landmarks

The occurrence of Q and J earlier than the echocardiographic landmark led to the underestimation of the QT and JT intervals by echocardiography and vice versa. In contrast, the occurrence of the end of the T-wave sooner than the echocardiographic landmark led to an overestimation of QT and JT. This should be considered when we measure the mechanical QT and JT intervals [Table 2].

Despite significant advances in cardiac resynchronization therapy in the pediatric population, the normal range of electromechanical dyssynchrony, defined as the time difference between the electrical event on the ECG and its mechanical counterpart on echocardiography, is not reported.[28],[29] However, this study shows the existence of dyssynchrony between the timing of ECG events and their M-mode echocardiographic counterparts in normal children.

A practical proposal for measuring mechanical QT and JT intervals

Based on our findings, we propose to measure QT and JT intervals by using an angled M-mode tracing of the aortic and mitral valve. The intervals are measured in five consecutive beats and are averaged [Figure 5]. This would yield a crude and reliable estimate of these intervals. However, a validation study is essential before this method can be applied to the fetus.

Future studies: Validation of mechanical QT and JT intervals

Since there is no study on mechanical QT and JT intervals, we have to compare our study with the available literature on mechanical PR intervals. Accordingly, two methods of validation have been described for mechanical PR interval.

Pasquini et al. compared electrical and mechanical PR intervals in 50 healthy fetuses and reported a mean difference of −16.47 ms with a 95% confidence interval of −43.43–10.44.[30] Similarly, Friedman et al. defined the agreement of prenatal echocardiographic and postnatal ECG measurements to validate the fetal mechanical PR interval.[31] Whereas, Glickstein et al. compared measurements of mechanical PR interval by 15 perinatologists or pediatric cardiologists and stated, a difference of equal or less than 30 msonds with the measurement of the principal reader is regarded as acceptable.[13] In other words, Pasquini and Friedman used agreement between prenatal mechanical measurements and postnatal electrical ECG as the validation method, while Glickstein et al. considered reproducibility as a surrogate of validation. Both of these methods are essential for the newly reported mechanical QT and JT intervals. Accordingly, serial measurement of fetal mechanical QT and JT intervals in a fetal cohort study and comparison with postnatal measurements are necessary as the initial step.

Future applications

This study sets the stage for future application of this technique to fetal echocardiography for timely diagnosis of the hereditary long QT syndromes in the prenatal period using anatomic M-mode echocardiography. Anatomic M-mode can overcome the technical challenges in the prenatal echocardiographic examination for obtaining appropriate images. The fetal ECG recording is time-consuming, and the equipment is not widely available in resource-limited countries.[22],[23],[24],[25] In contrast, the excellent temporal and axial resolution of M-mode echocardiography and display of the entire cardiac cycle in a single frame as a unidimensional imaging modality make it well suited for this purpose. Nevertheless, applying the findings of this study to fetal echocardiography is contentious and needs further investigation.

The main practical implication of this study is to provide the correlated echocardiographic points for the measurement of QT and JT intervals that have the potential to be used in fetal echocardiography. Furthermore, given the increasing popularity of the application of artificial intelligence and machine learning in echocardiography, another clinical application of this scientific observation may be the feasibility of automatic reporting QT and JT intervals and their corrected values during echocardiographic examination in the near future.[32]

   Conclusions Top

This is the first study on mechanical QT and JT intervals in healthy children. The study opens the horizons into the in utero diagnosis of CLQTS by measuring fetal QT and JT intervals using the widely available M-mode echocardiographic methods.

Limitations

In this study, we compared the mechanical QT and JT with concurrent ECG tracing on the monitor of the echocardiography machine. Therefore, the major limitation of this study was that we did not investigate the consistency between the mechanical QT and JT intervals with the surface ECG intervals. Thus, in the future, a comparative study for investigation of the reliability among these three sets of data on QT and JT intervals, including M-mode echocardiographic measurements, concurrently obtained ECG tracings on the monitor of the echocardiography machine and surface ECG (obtained immediately before or after the echocardiographic examination) is warranted. Furthermore, evaluating normal values of electromechanical window, as studied by Sugrue et al., is essential in healthy children.[33]

We did not use Color Doppler tissue M-mode for all of the measurements. The addition of Color Doppler tissue imaging might increase the precision of echocardiographic landmarks.

Ethics approval and consent to participate

We obtained informed consent from the children's parents or guardians and the participants, if applicable. We followed the principles stated in the Helsinki Declaration on ethical principles for medical research involving human subjects. The Institutional Committee approved this study for Ethics in Biomedical Research of the Medical School of Tehran University of Medical Sciences (ID: IR.TUMS.CHMC.REC.1399.190).[21]

Availability of data and materials

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Acknowledgment

The authors fully appreciate the kind permission of the distinguished colleagues Dr. Keyhan Sayadpour Zanjani, Dr. Ehsan Aghaei-Moghadam, Dr. Reza Shabanian, and Dr. Mojtaba Gorji to enroll their patients (referred to their outpatient clinic of the Children's Medical Center) in this study. The authors are grateful to the children and their parents for participation in this study.

Our special gratitude goes to the distinguished reviewers who helped us improve this manuscript's quality by their insightful and constructive comments.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 

   References Top
1.Patel R, Aggarwal S, Athwal PS, Randhawa S, Kahlon S. Long QT-syndrome with torsades de pointes managed considering financial constraints faced by the patient. Cureus 2021;13:e15892.  Back to cited text no. 1
    2.Leslie E, Medenbach A, Pittman E. Unmasking long QT syndrome in the emergency department: A case report. Clin Pract Cases Emerg Med 2021;5:89-92.  Back to cited text no. 2
    3.Asatryan B, Rieder M, Castiglione A, Odening KE. Arrhythmic risk during pregnancy and postpartum in patients with long QT syndrome. Herzschrittmacherther Elektrophysiol 2021;32:180-5.  Back to cited text no. 3
    4.Al-Akchar M, Siddique MS. Long QT syndrome. In: StatPearls. Treasure Island (FL): StatPearls Publishing Copyright © 2021, StatPearls Publishing LLC.; 2021.  Back to cited text no. 4
    5.Marwick TH. Echocardiography in long QT syndrome: The mechanical face of an electrical disease. J Am Coll Cardiol 2020;76:2844-6.  Back to cited text no. 5
    6.Heron SE, Hernandez M, Edwards C, Edkins E, Jansen FE, Scheffer IE, et al. Neonatal seizures and long QT syndrome: A cardiocerebral channelopathy? Epilepsia 2010;51:293-6.  Back to cited text no. 6
    7.Ishikawa S, Yamada T, Kuwata T, Morikawa M, Yamada T, Matsubara S, et al. Fetal presentation of long QT syndrome-evaluation of prenatal risk factors: A systematic review. Fetal Diagn Ther 2013;33:1-7.  Back to cited text no. 7
    8.Cherian AG, Lankala P, Krupa J, Roshan J. Ventricular noncompaction and long QT syndrome – A deadly double hit for the foetus. Indian Pacing Electrophysiol J 2021;21:186-90.  Back to cited text no. 8
    9.Verdurmen KM, Eijsvoogel NB, Lempersz C, Vullings R, Schroer C, van Laar JO, et al. A systematic review of prenatal screening for congenital heart disease by fetal electrocardiography. Int J Gynaecol Obstet 2016;135:129-34.  Back to cited text no. 9
    10.Yuan SM. Fetal arrhythmias: Surveillance and management. Hellenic J Cardiol 2019;60:72-81.  Back to cited text no. 10
    11.Wallace E, Howard L, Liu M, O'Brien T, Ward D, Shen S, et al. Long QT syndrome: Genetics and future perspective. Pediatr Cardiol 2019;40:1419-30.  Back to cited text no. 11
    12.Beckmann BM, Scheiper-Welling S, Wilde AA, Kääb S, Schulze-Bahr E, Kauferstein S. Clinical utility gene card for: Long-QT syndrome. Eur J Hum Genet 2021;29:1825-32.  Back to cited text no. 12
    13.Glickstein J, Buyon J, Kim M, Friedman D, PRIDE Investigators. The fetal Doppler mechanical PR interval: A validation study. Fetal Diagn Ther 2004;19:31-4.  Back to cited text no. 13
    14.Aghaei Moghadam E, Malakan Rad E, Nikoufar M, Ghamari A, Doosti-Irani A, Kocharian A, et al. Reference values of fetal mechanical PR interval and heart rate-corrected fetal mechanical PR interval: Influence of fetal sex, heart rate, gestational age, and maternal age. Int Cardio Res J 2018;12:e62358.  Back to cited text no. 14
    15.Rodríguez M, Moreno J, Márquez R, Eltit R, Martinez F, Sepúlveda-Martínez A, et al. Increased PR interval in fetuses of patients with intrahepatic cholestasis of pregnancy. Fetal Diagn Ther 2016;40:298-302.  Back to cited text no. 15
    16.Feigenbaum H. Role of M-mode technique in today's echocardiography. J Am Soc Echocardiogr 2010;23:240-57.  Back to cited text no. 16
    17.Zhou SH, Wong S, Rautaharju PM, Karnik N, Calhoun HP. Should the JT rather than the QT interval be used to detect prolongation of ventricular repolarization? An assessment in normal conduction and in ventricular conduction defects. J Electrocardiol 1992;25 Suppl: 131-6.  Back to cited text no. 17
    18.Salik J, Muskin PR. Consideration of the JT interval rather than the QT interval. Psychosomatics 2013;54:502.  Back to cited text no. 18
    19.Faul F, Erdfelder E, Lang AG, Buchner A. G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 2007;39:175-91.  Back to cited text no. 19
    20.Mitchell C, Rahko PS, Blauwet LA, Canaday B, Finstuen JA, Foster MC, et al. Guidelines for performing a comprehensive transthoracic echocardiographic examination in adults: Recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr 2019;32:1-64.  Back to cited text no. 20
    21.General Assembly of the World Medical Association. World Medical Association Declaration of Helsinki: Ethical principles for medical research involving human subjects. J Am Coll Dent 2014;81:14-8.  Back to cited text no. 21
    22.Matonia A, Jezewski J, Kupka T, Jezewski M, Horoba K, Wrobel J, et al. Fetal electrocardiograms, direct and abdominal with reference heartbeat annotations. Sci Data 2020;7:200.  Back to cited text no. 22
    23.Taylor MJ, Smith MJ, Thomas M, Green AR, Cheng F, Oseku-Afful S, et al. Non-invasive fetal electrocardiography in singleton and multiple pregnancies. BJOG 2003;110:668-78.  Back to cited text no. 23
    24.Desai L, Wakai R, Tsao S, Strasburger J, Gotteiner N, Patel A. Fetal diagnosis of KCNQ1-variant long QT syndrome using fetal echocardiography and magnetocardiography. Pacing Clin Electrophysiol 2020;43:430-3.  Back to cited text no. 24
    25.Donofrio MT, Moon-Grady AJ, Hornberger LK, Copel JA, Sklansky MS, Abuhamad A, et al. Diagnosis and treatment of fetal cardiac disease: A scientific statement from the American Heart Association. Circulation 2014;129:2183-242.  Back to cited text no. 25
    26.Tran PK, Tsang V. When and how to enlarge the small aortic root. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu 2016;19:55-8.  Back to cited text no. 26
    27.Scher AM, Young AC, Malmgren AL, Erickson RV. Activation of the interventricular septum. Circ Res 1955;3:56-64.  Back to cited text no. 27
    28.Ergul Y, Sevinc Sengul F, Öztürk E, Kafalı HC, Ayyıldız P, Tanıdır IC, et al. Cardiac resynchronization therapy in paediatric patients with congenital heart disease: Single centre with 10 years of experience. Cardiol Young 2021;31:940-8.  Back to cited text no. 28
    29.Friedberg MK, Mertens L. Echocardiographic assessment of ventricular synchrony in congenital and acquired heart disease in children. Echocardiography 2013;30:460-71.  Back to cited text no. 29
    30.Pasquini L, Seale AN, Belmar C, Oseku-Afful S, Thomas MJ, Taylor MJ, et al. PR interval: A comparison of electrical and mechanical methods in the fetus. Early Hum Dev 2007;83:231-7.  Back to cited text no. 30
    31.Friedman DM, Kim MY, Copel JA, Davis C, Phoon CK, Glickstein JS, et al. Utility of cardiac monitoring in fetuses at risk for congenital heart block: The PR Interval and Dexamethasone Evaluation (PRIDE) prospective study. Circulation 2008;117:485-93.  Back to cited text no. 31
    32.Kutty S. The 21st Annual Feigenbaum Lecture: Beyond artificial: Echocardiography from elegant images to analytic intelligence. J Am Soc Echocardiogr 2020;33:1163-71.  Back to cited text no. 32
    33.Sugrue A, van Zyl M, Enger N, Mancl K, Eidem BW, Oh JK, et al. Echocardiography-guided risk stratification for long QT syndrome. J Am Coll Cardiol 2020;76:2834-43.  Back to cited text no. 33
    

Top
Correspondence Address:
Prof. Elaheh Malakan Rad
Children's Medical Center (Pediatric Center of Excellence), Tehran University of Medical Sciences, No. 62, Dr. Gharib's Street, End of Keshavarz Boulevard, Tehran 1419733151
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None

Crossref citationsCheck

DOI: 10.4103/apc.apc_169_21

Rights and Permissions


  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
  [Table 1], [Table 2], [Table 3], [Table 4]